Non-chromatographic determination of ultratraces of V(V) and V(IV) based on a double column solid phase extraction flow injection system coupled to electrothermal atomic absorption spectrometry

In this work, a non-chromatographic procedure for the on-line determination of ultratraces of V(V) and V(IV) is presented. The method involves a solid phase extraction-flow injection system coupled to electrothermal atomic absorption spectrometry (SPE-FI-ETAAS). The system holds two microcolumns (MC...

Descripción completa

Detalles Bibliográficos
Autores: Kim, Manuela Leticia, Tudino, Mabel Beatriz
Tipo de recurso: artículo
Estado:Versión publicada
Fecha de publicación:2009
País:Argentina
Institución:Consejo Nacional de Investigaciones Científicas y Técnicas
Repositorio:CONICET Digital (CONICET)
Idioma:inglés
OAI Identifier:oai:ri.conicet.gov.ar:11336/72391
Acceso en línea:http://hdl.handle.net/11336/72391
Access Level:acceso abierto
Palabra clave:Electrothermal Atomic Absorption Spectrometry
Hybrid Mesoporous Silica
Solid Phase Extraction
https://purl.org/becyt/ford/1.4
https://purl.org/becyt/ford/1
Descripción
Sumario:In this work, a non-chromatographic procedure for the on-line determination of ultratraces of V(V) and V(IV) is presented. The method involves a solid phase extraction-flow injection system coupled to electrothermal atomic absorption spectrometry (SPE-FI-ETAAS). The system holds two microcolumns (MC) set in parallel and filled with lab-made mesoporous silica functionalized with 3-aminopropyltriethoxy silane (APS) and mesoporous silica MCM-41, respectively. The pre-concentration of V(V) is performed by sorption onto the first MC (C1) filled with APS at pH 3, whilst that of V(IV) is performed by sorption onto the second column (C2) filled with mesoporous silica MCM-41 at pH 5. Aqueous samples containing both analytes are loaded and, after pre-concentration (pre-concentration factor PCF = 10, sorption flow rate = 1 mL min-1, sorption time = 10 min), they are eluted in separate vessels with hydroxylammonium chloride (HC) 0.1 mol L-1 in HCl 0.5 mol L-1 (elution volume = 1 mL, elution flow rate = 0.5 mL min-1). Afterwards, both analytes are determined through ETAAS with graphite furnace. Under optimized conditions, the main analytical figures of merit for V(V) and V(IV) are, respectively: detection limits (3 s): 0.5 and 0.6 μg L-1, linear range: 2-100 μg L-1 (both analytes), sensitivity: 0.015 and 0.013 μg-1 L and sample throughput: 6 h-1 (both analytes). Recoveries of both species were assayed in different water samples. Validation was performed through certified reference materials for ultratraces of total vanadium in river water.