NaI(Tl) Scintillator's Response Functions for Point-like and Distributed Gamma-ray Sources

The response functions of a NaI(Tl) detectors have been estimated using Monte Carlo methods. Response functions were calculated for monoenergetic photon sources (0.05 to 3 MeV). Responses were calculated for point-like sources and for sources distributed in Portland cement cylinders. Calculated resp...

Descripción completa

Detalles Bibliográficos
Autores: Vega Carrillo, Héctor René, Benites Rengifo, Jorge Luís, Hernández Dávila, Víctor Martín, Ortíz Rodríguez, José Manuel, de León Martínez, Héctor Asael
Tipo de recurso: artículo
Estado:Versión publicada
Fecha de publicación:2016
País:México
Institución:Universidad Autónoma de Zacatecas
Repositorio:Repositorio Institucional Caxcán
Idioma:inglés
OAI Identifier:oai:http://ricaxcan.uaz.edu.mx:20.500.11845/785
Acceso en línea:http://localhost/xmlui/handle/20.500.11845/785
https://doi.org/10.48779/cysr-0875
Access Level:acceso abierto
Palabra clave:CIENCIAS FISICO MATEMATICAS Y CIENCIAS DE LA TIERRA [1]
Gamma- ray
Point-like source
Distributed source
NaI(Tl) response
Monte Carlo
Descripción
Sumario:The response functions of a NaI(Tl) detectors have been estimated using Monte Carlo methods. Response functions were calculated for monoenergetic photon sources (0.05 to 3 MeV). Responses were calculated for point-like sources and for sources distributed in Portland cement cylinders. Calculated responses were used to estimate the detector efficiency for pointlike and distributed sources. Samples of cylindrical Portland cement were prepared and exposed to the photoneutron field produced by a 15 MV linac used for radiotherapy. Short half-life radioisotopes were induced and the activity was determined by measuring the pulse-height spectra with a NaI(Tl) g-ray spectrometer that was calibrated using point-like sources. Instead of doing corrections due to differences between the geometry, material and solid angle of point-like sources used for calibration, and the Portland cement cylinders, the detection efficiency was determined using the ratio between the efficiencies for the point-like and the distributed sources estimated with the Monte Carlo calculations, and the activity of the induced isotopes in cement was obtained.